Comparison between Semi-Automated and Manual Methodologies for MTF Determination in a Digital Radiography System
DOI:
https://doi.org/10.29384/rbfm.2026.v20.19849001884Keywords:
Digital Radiology, Spatial Resolution, Quality Control, MTFAbstract
Spatial resolution, although commonly assessed qualitatively, can be quantified by determining the system's Modulation Transfer Function (MTF). The literature is extensive regarding methods for determining MTF in digital radiology equipment. Among the methodologies, there are semi-automatic methods, which require image analysis by dedicated software, and manual methods, which require the operator to position regions of interest (ROIs). In this study, three MTF measurement methods discussed in the literature – two semi-automatic and one manual – were compared using the RQA 5 beam quality of a Konica Altus DR digital radiology system in a large hospital. The semi-automatic methods showed strong agreement with each other, with variations between 0.01 and 5.49 MTF units (%), while the manual method showed greater divergence from the other two, varying between 0.04 and 6.14 MTF units (%), depending on the fine or broad focus. It is concluded that semi-automatic methods are advantageous in terms of reproducibility and efficiency when compared to the manual method. It is the medical physicist's responsibility to weigh the applicability of each methodology within the reality of their institution.
Downloads
References
1. Ministério da Saúde - MS, Agência Nacional de Vigilância Sanitária - ANVISA, Diretoria Colegiada. RESOLUÇÃO RDC No 611, DE 9 DE Março DE 2022. 2022 Mar 16;
2. Ministério da Saúde - MS, Agência Nacional de Vigilância Sanitária - ANVISA. INSTRUÇÃO NORMATIVA - IN N° 90, DE 27 DE MAIO DE
2021. Dispõe sobre requisitos sanitários para a garantia da qualidade e da segurança em sistemas de radiografia médica convencional, e dá outras providências. 2021. p. 6–10.
3. INTERNATIONAL STANDARD. ISO 15529:2010(E),Optics and photonics — Optical transfer function — Principles of measurement of modulation transfer function (MTF) of sampled imaging systems
4. Schneider CA, Rasband WS, Eliceiri KW. NIH Image to ImageJ: 25 years of image analysis. Nat Methods. 2012 Jul 28;9(7):671–5.
5. International Atomic Energy Agency - IAEA. IAEA HUMAN HEALTH SERIES NO. 39 -
Implementation of a Remote and Automated Quality Control Programme for Radiography and Mammography Equipment. 2021;
6. Gauntt D, Ranger N, Willis C, Al-Senan R, Bercha I, Dave J, et al. Acceptance Testing and Quality Control of Digital Radiographic Imaging Systems. 2024 Jul.
7. Coltman JW. The specification of imaging properties by response to a sine wave input. 1954.
8. Droege RT, Morin RL. A practical method to measure the MTF of CT scanners. 1982.
9. MicroDicom - free DICOM viewer for Windows [Internet]. [cited 2025 Feb 23]. Available from: https://www.microdicom.com/
10. Jerrold T. Bushberg. The Essential Physics of Medical Imaging. 3rd ed. 2011.
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2026 Jonas Nanni Menino, Tulio Guilherme Soares Marques, Raissa Alexia Camargo Guassu, Daniel Molena Seraphim, Diana Rodrigues de Pina

This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
The submission of original articles to the Brazilian Journal of Medical Physics implies the transfer, by the authors, of the rights of print and digital publication. Copyright for published articles remains with the author, with journal rights on first publication. Authors may only use the same results in other publications by clearly indicating this journal as the original publisher. As we are an open access journal, free use of articles in educational, scientific, non-commercial applications is allowed, as long as the source is cited.
The Brazilian Journal of Medical Physics is under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License (CC BY-NC-ND 4.0).